Capacitance Level Sensor: How It Works, Formula and Troubleshooting

Capacitance level sensor formula and calculator featured image

Capacitance Level Sensor: How It Works, Formula and Troubleshooting

WRITTEN BYKleev Technical Staff
PUBLISHED 
LAST UPDATED 
READ TIME12 min
VIEWS

A capacitance level sensor turns the probe and the tank wall into a simple capacitor, so the reading rises as liquid or powder covers the rod. Below is what a technician needs to size, calibrate or troubleshoot a capacitance level sensor without guessing.

Endress+Hauser’s capacitance measuring principle page puts it this way: the probe and the conductive tank wall form the capacitor, and a change in the medium between them changes the capacitance.

How does a capacitance level sensor work?

A capacitance level sensor measures the capacitance between an insulated or bare probe and the tank wall or a ground tube. Air has a dielectric constant near 1, while the process has a higher one, so capacitance climbs as the level rises. The electronics compare that value with the empty and full readings set at calibration and send a relay change or a 4 to 20 mA signal.

What is a capacitance level sensor and where does the signal come from?

A capacitance level sensor is a probe that reads level as a change in capacitance, measured in picofarads (pF). The signal comes from the material around the rod, because every material stores more electric field than air does.

The electronics drive a small radio frequency signal onto the probe and measure the current that returns through the vessel wall. The level sensor entry on Wikipedia notes that this is why the method is often called RF level, and that it works on liquids, slurries and bulk solids.

In a capacitance level sensor the rod is one plate, the tank wall or a ground tube is the other, and the vapor, liquid or powder between them is the dielectric. As the process covers more rod, the average dielectric rises and so does the capacitance level sensor reading.

Capacitance level sensor probe in a ground tube with oil level and pF values
Capacitance level sensor with a 0.5 in rod in a 2 in ground tube: 61.2 pF empty and 105.2 pF at 36 in of oil.

What is the formula for a capacitance level sensor?

For a rod centered in a round tube or tank, capacitance equals 2π times ε0 times εr times the covered length, divided by the natural log of D over d. This is the textbook coaxial capacitor equation applied to a capacitance level sensor.

C = 2π × ε0 × εr × L / ln(D/d)

Where: C = capacitance, F; ε0 = 8.854 × 10^-12 F/m; εr = dielectric constant of the material; L = covered probe length, m; D = inside diameter of tube or tank; d = rod diameter.

Shop form: C per inch = 1.413 × εr / ln(D/d), pF/in

The Engineering ToolBox permittivity table lists the vacuum permittivity ε0 as 8.854 × 10^-12 F/m, which works out to 1.413 pF per inch of length before you apply the geometry term.

A capacitance level sensor probe is really two capacitors stacked along its length, wet below and dry above:

C(h) = [1.413 / ln(D/d)] × [εr × h + 1.0 × (L − h)], pF

Where: h = level above the probe tip, in; L = active probe length, in.

That relation is linear in h, which is why a capacitance level sensor needs only two calibration points, empty and full.

How do you calculate capacitance level sensor output? A solved example

Work out the pF per inch from the geometry, then multiply by the weighted dielectric over the probe length. The capacitance level sensor in this example is a bare rod in a ground tube on a lube oil tank.

Solved example

A 0.5 in (12.7 mm) rod sits inside a 2 in (50.8 mm) ground tube, with an active length of 60 in (1,524 mm). The oil has εr = 2.2, the Engineering ToolBox value for mineral transformer oil, and the level is 36 in (914 mm).

  1. Geometry term: ln(2 / 0.5) = ln 4 = 1.386, so the capacitance level sensor gives 1.413 / 1.386 = 1.019 pF per inch (40.1 pF/m) in air.
  2. Empty: 1.019 × 60 = 61.2 pF.
  3. At 36 in: 1.019 × (2.2 × 36 + 1.0 × 24) = 1.019 × 103.2 = 105.2 pF.
  4. Full: 1.019 × 2.2 × 60 = 134.6 pF, so the calibrated span is 134.6 − 61.2 = 73.4 pF.
  5. Check: (105.2 − 61.2) / 73.4 = 60 percent, the same as 36 / 60 in, so the output is 13.6 mA on a 4 to 20 mA loop.

DwyerOmega’s application notes give a minimum span of about 4 pF and an upper range limit near 2,500 pF for a typical two wire capacitance level sensor transmitter. The 73.4 pF span above sits well inside that window.

Capacitance level sensor calculator

Empty 61.2 pF, at level 105.2 pF, full 134.6 pF, span 73.4 pF

The calculator assumes a nonconductive process and a capacitance level sensor rod centered in a round metal tube. With the rod off center in a large tank, treat the result as an estimate.

How does dielectric constant affect a capacitance level sensor?

The higher the dielectric constant, the bigger the change in pF per inch of level on a capacitance level sensor, and materials close to 1.0 give tiny spans.

Wikipedia’s level sensor page lists dielectric constants from about 1.1 for coke and fly ash up to 88 for water. Engineering ToolBox gives 80 for water at 68 °F (20 °C), 2.2 for mineral transformer oil and polypropylene, and 2.5 to 3.5 for silica sand.

On the solved example’s capacitance level sensor, plastic pellets at εr 1.1 give a full span of only 6.1 pF, while oil at 2.2 gives 73.4 pF. Water would give thousands of pF on a bare rod, but its conductivity changes the picture.

Insulated vs bare probe: which capacitance level sensor fits your liquid?

Use a bare capacitance level sensor for nonconductive liquids and dry solids, and an insulated one for conductive, water based liquids. Conductivity decides it, not dielectric constant alone.

Endress+Hauser places conductive liquids above about 100 µS/cm and nonconductive ones such as oils and solvents below about 1 µS/cm. With a conductive liquid, the liquid itself becomes the outer plate, and the probe insulation becomes the dielectric.

On an insulated probe in a conductive liquid, the capacitance level sensor reading no longer depends on the liquid’s dielectric constant or on the tank shape, so factory calibration is often possible.

Between those two sits the trouble zone: Endress+Hauser warns that from 1 to 100 µS/cm a small change in conductivity can make the reading jump, so avoid that range for a capacitance level sensor.

A plastic or fiberglass vessel needs a return path. A second rod, a ground tube or a conductive strip down the outside wall gives the capacitance level sensor its second plate.

Insulation also sets the capacitance level sensor temperature limit. DwyerOmega lists nylon and Ryton insulators at about 350 to 450 °F (175 to 230 °C), with ceramic used above that.

Point vs continuous: what is the difference?

A point capacitance level sensor switches a relay at one height, while a continuous one sends a signal proportional to level across the whole probe. Only the electronics differ.

A point capacitance level sensor usually mounts through the side wall with a short rod. Kleev level switches cover high, low and pump control duties of this kind.

A continuous capacitance level sensor hangs vertically and sends 4 to 20 mA, often with HART. If you need a refresher on loop signals, this guide to the 4 to 20 mA current loop covers the basics.

How do you calibrate a capacitance level sensor at empty and full?

Store the empty value with the probe in vapor and the full value with it covered; the transmitter scales between them. For a nonconductive process, take both points with the real product.

  1. Confirm the capacitance level sensor type matches the liquid conductivity and the ground path is solid.
  2. With the level below the probe tip, store the empty value (61.2 pF in the solved example).
  3. Fill to the top of the active length, or a known level, and store that value.
  4. At a partial level, the linear relation does the rest: 105.2 pF at 36 in of oil implies 134.6 pF full.
  5. Check one middle point against a sight glass or dip tape and record both pF values.

Use a documenting loop calibrator to confirm the capacitance level sensor mA output at each point. Recalibrate each time the product changes.

Troubleshooting flowchart for a capacitive level probe reading wrong
Capacitance level sensor troubleshooting flowchart: check the ground, the conductivity range, coating and calibration in that order.

Why is my capacitance level sensor reading wrong?

Most bad readings come from four things: coating on the probe, a lost ground path, a change in the product dielectric, or calibration taken on the wrong product. Check them in that order before you replace the electronics.

SymptomLikely causeFix
Reading stays high after drainingCoating buildup on the rod or insulationClean the probe; choose a build up immune design with a driven shield
Reading jumps or drifts with no level changeConductivity in the 1 to 100 µS/cm zoneChange to an insulated probe or a different technology
Span much smaller than expectedLow dielectric product such as pellets or fly ashUse a ground tube or smaller gap to raise pF per inch
Zero shifts after maintenancePoor ground between housing and tank, or loose coaxBond the housing to the vessel and check the cable to the remote head
Reads full in an empty plastic tankNo second plate on a nonconductive vesselAdd a ground rod, ground tube or conductive strip
Wrong level after a product changeNew dielectric value, old calibrationRecalibrate empty and full with the new product
Noisy signal with agitationFoam, splashing or turbulenceFit a stilling well and add damping

On coating, DwyerOmega explains that a driven shield runs at the same potential and frequency as the measuring probe, so no current flows through the coating and the capacitance level sensor ignores it.

In dry solids, Wikipedia notes that moving low dielectric powders can build a high voltage charge, so ground every capacitance level sensor properly.

What are the pros and cons of a capacitance level sensor?

A capacitance level sensor is simple, has no moving parts and handles liquids, slurries and powders. Its weak points are coating, changing dielectric values and the need for a good ground path.

Pros

  • No moving parts, so nothing to stick or wear
  • Works on liquids, slurries and bulk solids
  • Low cost compared with radar
  • Insulated probes handle many acids and caustics

Cons

  • Coating can read as level unless the probe has a driven shield
  • Nonconductive products need calibration with the real product
  • Small spans on products near εr 1.1
  • Unstable in the 1 to 100 µS/cm conductivity range

If you compare technologies, our posts on radar level transmitters and magnetic level gauges cover the noncontact and float based alternatives to a capacitance level sensor. The hydrostatic level method is often the better fit for open water basins.

Myth vs fact: capacitance level sensor edition

Several shop floor beliefs about the capacitance level sensor are only half true, and these are the ones that come up most during commissioning.

MYTH

A higher dielectric constant always gives a better capacitance level sensor reading.

FACT

Once a liquid is conductive, the insulation sets the capacitance and the dielectric of the liquid stops mattering.

MYTH

Any capacitance level sensor works out of the box on any tank.

FACT

Nonconductive products need an empty and full calibration with the real product, because span depends on εr and tank geometry.

MYTH

Coating on a capacitance level sensor only matters if it is thick.

FACT

A thin conductive film on an insulated rod can hold the reading near full; a driven shield is the real cure.

MYTH

Plastic tanks cannot use a capacitance level sensor.

FACT

They can, once you add a ground rod, a ground tube or a conductive strip as the second plate.

Industrial applications of the capacitance level sensor

A capacitance level sensor fits wherever a simple level signal is enough and the product is stable. These four cases are typical in US plants.

Refinery

Insulated probes in ground tubes track lube oil and seal oil reservoirs, where εr near 2.2 gives a clean, repeatable span. A point capacitance level sensor on a slop drum gives the high alarm.

Water and wastewater plant

Insulated rods measure chemical feed tanks such as sodium hypochlorite and caustic, where conductivity makes the reading independent of dielectric. Wikipedia gives sewage slurry a dielectric near 50.

HVAC

A point capacitance level sensor guards condensate pans, cooling tower basins and humidifier reservoirs against overflow or dry running.

Chemical plant

PP or PFA coated probes handle acids and alkalis, and Endress+Hauser rates its two rod Liquicap FMI21 to 145 psi (10 bar) and 212 °F (100 °C). Cable probes track plastic pellet silos with a ground tube or tight gap.

To see where this method sits among the other options, read this overview of level transmitter types. Kleev’s range of level transmitters and level gauges covers the same duties with other principles.

Coaxial capacitance formula card with dielectric constants and pF spans
Capacitance level sensor formula card with full span for pellets, oil and water on the same 60 in probe.

Frequently asked questions

Can a capacitance level sensor measure oil and other nonconductive liquids?

Yes. Oils and solvents below about 1 µS/cm suit a bare rod or a probe in a ground tube, and with transformer oil at εr 2.2 our 60 in (1,524 mm) capacitance level sensor gives a 73.4 pF span. Calibrate empty and full on the actual oil, not on water.

What is the dielectric constant of water for level measurement?

Engineering ToolBox gives about 80 at 68 °F (20 °C), and Wikipedia’s level sensor page quotes up to 88. In practice tap and process water is also conductive, so you use an insulated probe. The reading then depends on the insulation, not on that high dielectric value.

Why does coating cause errors on a capacitive probe?

Sticky or conductive film stays on the rod after the level falls, so the electronics still see a covered probe and read high. A probe with a driven shield holds the coating at the same potential as the rod, so no current flows through it. Cleaning schedules help on probes without that feature.

How long can a capacitance level probe be?

Rigid rods typically run a few feet, and Endress+Hauser builds its two rod FMI21 up to 8 ft (2.5 m). For silos, Wikipedia cites cable probes of 20 to 50 m (about 66 to 164 ft). Long cables need anchoring and a check of the tension load from the powder.

Does a capacitance level sensor need recalibration when the product changes?

For a nonconductive process, yes, because the span scales with the dielectric constant. Swap a 2.2 oil for a 1.1 pellet product and the span on our capacitance level sensor falls from 73.4 pF to 6.1 pF. Conductive liquids on an insulated probe usually keep their calibration.

Can a capacitive level probe work in a plastic tank?

Yes, but the plastic wall cannot act as the second plate. Add a ground tube around the rod, a second parallel rod or a conductive strip bonded down the outside wall. The ground tube option also gives a defined geometry, so you can predict pF per inch with the coaxial formula.