RTD Calibration: Step by Step Procedure & Pt100 Tolerance Calculator
Table of Contents
- How do you do an RTD calibration on a Pt100?
- What is RTD calibration and why does a Pt100 drift?
- What equipment do you need for RTD calibration?
- How do you calibrate an RTD step by step?
- RTD calibration step 1: inspect and check insulation
- RTD calibration step 2: ice point check
- RTD calibration step 3: load the dry block or bath
- RTD calibration step 4: stabilize and record
- RTD calibration step 5: convert and compare
- What is the RTD calibration formula for Pt100 resistance?
- How do you calculate RTD calibration error? A solved example
- Solved example
- Pt100 tolerance calculator
- How do 2, 3 and 4 wire connections affect RTD calibration?
- Why is my RTD calibration out of tolerance?
- How often should RTD calibration be done?
- Pros and cons of RTD calibration methods
- Pros
- Cons
- Myth vs fact: RTD calibration
- Industrial applications of RTD calibration
- Refinery
- Water and wastewater plant
- HVAC
- Chemical plant
- Frequently asked questions
- What is the acceptable tolerance for RTD calibration on a Class A Pt100?
- Can you do an RTD calibration with only an ice bath?
- Why use a 4 wire connection for RTD calibration of a Pt100?
- What is the resistance of a Pt100 at 100 degrees Celsius?
- What immersion depth does RTD calibration in a dry block need?
- Can a probe that fails RTD calibration be adjusted or replaced?
RTD calibration tells you how far a Pt100 probe has drifted from the IEC 60751 curve, and whether it still sits inside its Class A or Class B tolerance. For technicians and engineers, below is a repeatable bench procedure, a solved example and a calculator that does the Callendar Van Dusen math for you.
A Pt100 reads 100 Ω at 32 °F (0 °C) and changes by roughly 0.385 Ω per °C, according to Beamex’s Pt100 sensor guide. That small slope is why lead wire, immersion depth and a shaky reference can swamp an RTD calibration.
How do you do an RTD calibration on a Pt100?
RTD calibration on a Pt100 compares the probe’s resistance with a traceable reference thermometer at several stable temperatures, then converts each reading to temperature with the Callendar Van Dusen equation. You pass the probe when every error stays inside its IEC 60751 class limit, for example ±0.35 °C at 100 °C for Class A.
- Inspect the probe and check insulation resistance.
- Run an ice point check at 32 °F (0 °C).
- Insert the probe and a reference PRT in a dry block or bath.
- Read 4 wire resistance at 3 or more stable points.
- Convert to °C, compare with the class limit, and record the RTD calibration result.
What is RTD calibration and why does a Pt100 drift?
RTD calibration is a measured comparison between your sensor and a better reference at known temperatures. It does not adjust the platinum element itself; it tells you the error so you can accept, correct or replace the probe.
Platinum is stable, but between each RTD calibration the element still shifts with thermal cycling, vibration, moisture in the sheath and strain from rough handling. Endress+Hauser’s white paper on temperature calibration notes that sensors deviate over time because of drift and aging, and that the risk of an undetected out of spec reading grows the longer you go between checks.
If you are still choosing a sensor type, our comparison of RTDs and thermocouples covers the tradeoffs. For thermocouple work, follow the separate thermocouple calibration procedure, since the reference junction and ASTM E230 limits change the method.
What equipment do you need for RTD calibration?
You need a stable heat source, a reference thermometer more accurate than the probe, and a readout that measures resistance in 4 wire mode. An ice point bath is the cheapest useful extra on an RTD calibration bench.
- Dry block or stirred bath covering your RTD calibration range.
- Reference PRT with a current calibration certificate and its own readout.
- RTD readout or precision ohmmeter with 4 wire connection and low excitation current.
- Ice point flask, distilled water and a blender for shaved ice.
- Megohmmeter for the sheath insulation check.
Fluke’s dry block versus bath guide says dry blocks suit straight, small diameter probes, while baths suit long, odd shaped sensors or RTDs with a transmitter head. Baths also give better stability, because stirred fluid transfers heat better than a metal well.
If the probe feeds a head transmitter, check the output side as well, and this guide to the mA signal loop covers that end.
A field temperature calibrator can cover the dry block and readout roles in one case.

How do you calibrate an RTD step by step?
Start with inspection and the ice point, then work up through your comparison points and let each one stabilize fully. The RTD calibration flowchart below shows the same order.
RTD calibration step 1: inspect and check insulation
Look for a bent sheath, crushed cable and corroded terminals before you heat anything. Then measure insulation resistance between the leads and the sheath, because moisture inside the probe creates a shunt that pulls the reading low; this guide to insulation resistance testing covers safe megger practice.
RTD calibration step 2: ice point check
Measurement Canada’s ice point procedure uses shaved ice from distilled water packed in a vacuum flask, drained of free water and left to settle for 15 minutes. Immerse the probe at least 2.8 inches (7 cm) below the surface and keep it 1.2 to 1.6 inches (3 to 4 cm) above the bottom.
Done this way, the same document puts the ice point uncertainty near 0.01 °C. Beamex recommends repeating the 0 °C point during the run, since a shift in R0 is the clearest sign the probe moved.
RTD calibration step 3: load the dry block or bath
Place the reference PRT and the probe under test in matching wells, with both sensing zones level at the bottom. Fluke’s guidance puts minimum immersion at 15 times the sensor diameter plus the element length, so a 0.25 inch (6.35 mm) probe needs about 3.75 inches (95 mm) plus its element.
RTD calibration step 4: stabilize and record
Set the first point, wait for both readouts to settle, and log several readings rather than one. Fluke’s dry well article uses a 30 minute window to characterize block stability, which tells you not to rush the RTD calibration either.
RTD calibration step 5: convert and compare
Turn each measured resistance into temperature with the Callendar Van Dusen equation, subtract the reference value, and compare the error to the class tolerance. Repeat the 0 °C point at the end to check for hysteresis.
What is the RTD calibration formula for Pt100 resistance?
Above 0 °C the Pt100 follows R(t) = R0 (1 + A t + B t²), and Class A tolerance is ±(0.15 + 0.002 |t|) °C. Below 0 °C a third term, C (t − 100) t³, is added.
R(t) = R0 × (1 + A×t + B×t²) for t ≥ 0 °C
R(t) = R0 × (1 + A×t + B×t² + C×(t − 100)×t³) for t < 0 °C
A = 3.9083E−3, B = −5.775E−7, C = −4.183E−12, R0 = 100 Ω
Class A: ±(0.15 + 0.002|t|) °C Class B: ±(0.30 + 0.005|t|) °C
The coefficients are the IEC 60751 values for the 385 curve listed in NI’s RTD documentation. The history of the Callendar and Van Dusen equation explains that Van Dusen, at the National Bureau of Standards (now NIST), added the low temperature term in 1925.
The tolerance formulas come from WIKA’s summary of IEC 60751 classes, which also lists Class AA at ±(0.1 + 0.0017 |t|) °C and Class C at ±(0.6 + 0.01 |t|) °C. WIKA gives Class A validity as -148 to 842 °F (-100 to 450 °C) for wire wound elements and -22 to 572 °F (-30 to 300 °C) for thin film.
How do you calculate RTD calibration error? A solved example
Convert the measured ohms to temperature, subtract the reference temperature, and compare the result with the class limit at that point. Here is a three point RTD calibration on a Class A probe.
Solved example
A Class A Pt100, 4 wire, is checked at 32 °F (0 °C), 212 °F (100 °C) and 392 °F (200 °C). The readout shows 100.04 Ω, 138.62 Ω and 175.95 Ω.
- Ideal resistance at 212 °F: R = 100 × (1 + 0.39083 − 0.005775) = 138.51 Ω.
- 138.62 Ω inverted with Callendar Van Dusen gives 100.30 °C, so the error is +0.30 °C (+0.54 °F).
- Class A limit at 100 °C: 0.15 + 0.002 × 100 = ±0.35 °C (±0.63 °F), so this point passes.
- At 32 °F, 100.04 Ω gives +0.10 °C (+0.18 °F) against a ±0.15 °C (±0.27 °F) limit: pass.
- At 392 °F, 175.95 Ω against an ideal 175.86 Ω gives +0.26 °C (+0.46 °F) against ±0.55 °C (±0.99 °F): pass.
Result: the probe passes RTD calibration, but the 212 °F point uses 86% of its band, so shorten the next interval.
Pt100 tolerance calculator
How do 2, 3 and 4 wire connections affect RTD calibration?
A 2 wire circuit adds both lead resistances to the reading, a 3 wire circuit cancels them if the leads match, and a 4 wire circuit removes them almost completely. For any RTD calibration on the bench, use 4 wire measurement.
Copper wire resistance tables from Engineering ToolBox give 22 AWG at 16.5 Ω per 1,000 ft at 77 °F (25 °C). Fifty feet (15.2 m) of 2 wire cable therefore adds about 1.65 Ω, which a Pt100 reads as roughly 7.7 °F (4.3 °C) high.
That single error is more than twelve times the Class A band at 212 °F. If you want to try other lengths and gauges, the lead wire error calculator for Pt100 and Pt1000 covers 2 and 3 wire cases.
Beamex notes that IEC 60751 requires 3 or 4 wire measurement for any sensor better than Class B. A 3 wire probe still carries error equal to the mismatch between its leads, so record the wiring type on every RTD calibration certificate.
Why is my RTD calibration out of tolerance?
Most failed RTD calibration results come from the setup, not the sensor: short immersion, an unstable source, the wrong curve or uncompensated leads. Rule those out before you condemn the probe.
| Symptom | Likely cause | Fix |
|---|---|---|
| Reads high at every point by a fixed amount | Lead resistance in a 2 wire hookup | Repeat the RTD calibration in 4 wire |
| Error grows with temperature | Wrong curve set, such as 391 instead of 385 | Match alpha on the readout to the probe nameplate |
| Reads low at high points only | Stem conduction from short immersion | Insert deeper; use a bath for short probes |
| Reading wanders and will not settle | Block not stable or loose well fit | Use the correct sleeve and restart the RTD calibration |
| Reads low, worse when hot | Moisture lowering insulation resistance | Megger test; dry or replace the probe |
| R0 shifts after the run | Strain or damage in the element | Repeat the ice point; fail the RTD calibration if it stays off |
Beamex gives a good example of the curve mistake: mixing up 385 and 391 sensors causes about 1.5 °C error at 100 °C. Self heating is the other quiet cause, so keep excitation current at 1 mA or less, as the same guide recommends.
Ambient swings around the readout and cable also matter, a point we cover in how ambient conditions affect instrument accuracy.

How often should RTD calibration be done?
Most plants start RTD calibration at a fixed interval and then adjust it from the history. Endress+Hauser describes a typical scheduled interval of six months, following manufacturer advice.
Shorten the interval when a probe uses most of its band, as in our example, or when it sees heavy thermal cycling. Lengthen it after several clean results in a row, and log every RTD calibration so the trend is visible.
Remember that a loop check is not a sensor check. If the probe feeds a temperature transmitter, calibrate the sensor and the transmitter separately, or as a matched pair with the sensor's own coefficients loaded.
Pros and cons of RTD calibration methods
Dry block RTD calibration by comparison is fast and portable, while a bath and fixed point checks give lower uncertainty. Most shops use both.
Pros
- Dry block RTD calibration runs on site with no fluids.
- A stirred bath handles long probes and transmitter heads.
- Ice point RTD calibration needs little gear and catches R0 drift.
- Comparison with a reference PRT beats the block display.
Cons
- Dry blocks suffer from loose well fit and loading errors.
- Baths are heavy, slow to heat and need fluid care.
- An ice point RTD calibration covers only one temperature.
- Probes installed in a thermowell cannot be checked in place.
Myth vs fact: RTD calibration
A few shop habits cause more failed RTD calibration results than real drift does.
A Pt100 that reads 100 Ω in ice needs no further RTD calibration.
One point says nothing about slope. Check at least three points spread over the working range.
The dry block display is a good enough RTD calibration reference.
Fluke recommends a separate reference PRT for better accuracy, because the block's control sensor sits away from your probe.
Lead wire error is too small to matter.
Fifty feet (15.2 m) of 22 AWG in 2 wire adds about 7.7 °F (4.3 °C) to the reading.
Class A applies at any temperature.
WIKA lists Class A for thin film only from -22 to 572 °F (-30 to 300 °C).
Industrial applications of RTD calibration
Any loop where a few degrees change product quality, energy cost or a safety trip needs scheduled RTD calibration.
Refinery
Crude unit heater outlet and reactor bed RTDs feed control and trip logic. Regular RTD calibration stops a drifted probe from pushing a furnace hotter than needed.
Water and wastewater plant
Digester temperature drives biogas yield, so operators run RTD calibration at the ice point and one warm point.
HVAC
Chilled water supply and return RTDs set chiller staging and energy billing. Matched probe pairs need RTD calibration together, since a small offset skews the delta T.
Chemical plant
Batch reactor jackets and sterilization steps use 3 wire Pt100s; Endress+Hauser describes SIP holds near 250 °F (121 °C) for up to an hour.
Probes in these services usually sit in a thermowell, so you pull the insert for the bench check without opening the process. Kleev supplies replacement RTD and thermocouple assemblies when a probe fails its RTD calibration.

Frequently asked questions
What is the acceptable tolerance for RTD calibration on a Class A Pt100?
IEC 60751 sets Class A at ±(0.15 + 0.002 |t|) °C, which is ±0.15 °C (±0.27 °F) at the ice point, ±0.35 °C (±0.63 °F) at 212 °F and ±0.55 °C (±0.99 °F) at 392 °F. Your RTD calibration passes when every point sits inside these limits.
Can you do an RTD calibration with only an ice bath?
An ice bath gives one very good point, near 0.01 °C uncertainty when prepared carefully, and it catches shifts in R0. It cannot show slope errors, so add at least two higher points in a dry block or bath against a reference thermometer before you sign the certificate.
Why use a 4 wire connection for RTD calibration of a Pt100?
Four wire measurement sends current through one pair and senses voltage on the other, so lead resistance drops out. With 2 wire, 50 ft (15.2 m) of 22 AWG adds about 1.65 Ω, or 7.7 °F (4.3 °C), which swamps a Class A band of ±0.63 °F at 212 °F.
What is the resistance of a Pt100 at 100 degrees Celsius?
An ideal IEC 60751 Pt100 reads 138.51 Ω at 212 °F (100 °C), using A = 3.9083E−3 and B = −5.775E−7 in the Callendar Van Dusen equation. At 392 °F (200 °C) it reads 175.86 Ω, and at 32 °F (0 °C) exactly 100 Ω.
What immersion depth does RTD calibration in a dry block need?
Fluke gives a minimum of 15 times the sensor diameter plus the length of the sensing element, so a 0.25 inch (6.35 mm) probe needs about 3.75 inches (95 mm) plus its element. If the well is too shallow, stem conduction pulls the reading toward ambient, so use a bath.
Can a probe that fails RTD calibration be adjusted or replaced?
The platinum element cannot be trimmed, but you can load the measured Callendar Van Dusen coefficients into a transmitter that accepts them, or apply an offset in the control system. If R0 keeps shifting or insulation resistance is low, replace the probe instead of correcting it.