Pressure Gauge Accuracy: ASME B40.100 Grades & Error Calculator
Table of Contents
- What does pressure gauge accuracy mean on a dial gauge?
- How is pressure gauge accuracy calculated?
- What are the ASME B40.100 pressure gauge accuracy grades?
- What does 1% accuracy mean on a pressure gauge?
- What is the difference between percent of span and percent of reading?
- How do you work out the pressure gauge accuracy band? A solved example
- Solved example: pressure gauge accuracy at 120 psi
- Pressure gauge accuracy calculator
- Pressure gauge accuracy band calculator
- What range should a pressure gauge be?
- Range rules for pressure gauge accuracy
- Why does a pressure gauge lose accuracy?
- How do you troubleshoot poor pressure gauge accuracy?
- Field tips for keeping pressure gauge accuracy
- Pros and cons of higher pressure gauge accuracy grades
- Pros
- Cons
- Myth vs fact: pressure gauge accuracy
- Industrial applications of pressure gauge accuracy grades
- Refinery
- Water and wastewater plant
- HVAC
- Chemical plant
- How often should you check pressure gauge accuracy?
- Frequently asked questions
- What is the standard for pressure gauge accuracy in the US?
- Is ASME Grade A or Grade 1A more accurate?
- How do I convert a pressure gauge accuracy grade into psi?
- Why should normal pressure sit near the middle of the gauge scale?
- Does temperature change pressure gauge accuracy on a dial gauge?
- Can a liquid filled gauge help readings on a pulsing pump?
Pressure gauge accuracy tells you how far a dial reading can sit from the true pressure, and this page shows how to turn a grade into a psi band. It’s for the technicians, engineers and buyers who pick the grade and range, then keep the gauge in tolerance.
ASME B40.100 sets the pressure gauge accuracy grades used on most US dial gauges, from Grade 4A at 0.1% of span to Grade B at 3% of span near the ends of the dial. Once you know the grade, the span and where normal pressure sits, the math takes one line.
What does pressure gauge accuracy mean on a dial gauge?
Pressure gauge accuracy is the largest error allowed at any point on the dial, stated as a percent of span under ASME B40.100. A Grade A gauge on a 0 to 200 psi (0 to 13.8 bar) range may read within 2.0 psi (0.14 bar) in the middle half and within 4.0 psi (0.28 bar) near the ends. Pick a range that puts normal pressure near mid scale for the best pressure gauge accuracy.
How is pressure gauge accuracy calculated?
Multiply the grade percentage by the span, which is full scale minus the low end of the range. The answer is the pressure gauge accuracy band, a plus or minus value in psi that stays the same size anywhere inside that grade zone.
WIKA gives a simple case on its pressure gauge basics page: a 0 to 100 psi gauge rated at 1.0% reads between 49 and 51 psi when the true pressure is 50 psi. That 1 psi (0.069 bar) band holds at 10 psi as well, so the error as a share of the reading grows as pressure drops.
Pressure gauge accuracy band = ± (Grade % / 100) × (Full scale − Low end)
Percent of reading = Band / Reading × 100
Grades A and B: use the middle half value only from 25% to 75% of span.
This is why pressure gauge accuracy is quoted as percent of span on almost every dial gauge datasheet.
What are the ASME B40.100 pressure gauge accuracy grades?
ASME B40.100 defines Grades 4A, 3A, 2A, 1A, A and B, plus coarser grades, each tied to a percent of span. Grades 4A to 1A hold one pressure gauge accuracy value across the dial, while Grades A and B allow a wider band in the first and last quarter.
The current edition is ASME B40.100 2022, “Pressure Gauges and Gauge Attachments”, and its table of contents on the ASME site lists the grades and recommended test points in Table 2-2.4-1.
| Grade | Lower quarter | Middle half | Upper quarter | Typical use |
|---|---|---|---|---|
| 4A | ± 0.1% | ± 0.1% | ± 0.1% | Test and reference gauges |
| 3A | ± 0.25% | ± 0.25% | ± 0.25% | Test gauges, critical service |
| 2A | ± 0.5% | ± 0.5% | ± 0.5% | Tight pressure gauge accuracy for control |
| 1A | ± 1% | ± 1% | ± 1% | General process gauges |
| A | ± 2% | ± 1% | ± 2% | Industrial service gauges |
| B | ± 3% | ± 2% | ± 3% | Commercial and utility gauges |
Read every row as pressure gauge accuracy in percent of span. A Grade 1A gauge on 0 to 300 psi (0 to 20.7 bar) is good to 3 psi (0.21 bar) everywhere, while Grade A on the same range is good to 3 psi mid scale and 6 psi (0.41 bar) at the ends.

What does 1% accuracy mean on a pressure gauge?
It means the reading can be off by 1% of the full span at any point on the dial. On a 0 to 100 psi gauge that pressure gauge accuracy band is 1 psi (0.069 bar), whether the pointer sits at 90 psi or at 10 psi.
Processing magazine shows the effect: a 1% of full scale gauge is 1% of actual at 100 units, 2% at 50, 4% at 25 and 10% at 10. So pressure gauge accuracy at the low end of the dial is poor in relative terms, however good the grade looks.
What is the difference between percent of span and percent of reading?
Percent of span uses one fixed band for the whole range, and percent of reading shrinks the band as pressure falls. Most dial gauges state pressure gauge accuracy as percent of span, while some digital test gauges quote percent of reading.
Instrumart puts it in numbers: a 0.1% of reading gauge is within 0.1 psi at 100 psi and within 0.05 psi at 50 psi. Processing magazine adds that some percent of reading gauges only hold that spec from 25% to 100% of full scale, so read the datasheet before you compare pressure gauge accuracy across two types.
How do you work out the pressure gauge accuracy band? A solved example
Find the span and the zone the reading sits in, multiply by the grade, then add any temperature error from the datasheet. Here it is for a Grade A gauge on a pump discharge header.
Solved example: pressure gauge accuracy at 120 psi
A 0 to 200 psi (0 to 13.8 bar) Grade A gauge reads 120 psi (8.27 bar) in a pump room at 104 °F (40 °C). The datasheet temperature error is 0.4% of span per 18 °F (10 K) away from 68 °F (20 °C), the figure WIKA lists for its 332.54 gauge.
- Span = 200 − 0 = 200 psi (13.8 bar). The reading is at 120 / 200 = 60% of span.
- 60% is in the middle half, so Grade A pressure gauge accuracy is ± 1% of span = ± 2.0 psi (0.14 bar).
- Temperature offset = 104 − 68 = 36 °F, or 2 steps of 18 °F. Extra error = 2 × 0.4% × 200 = ± 1.6 psi (0.11 bar).
- Worst case total = 2.0 + 1.6 = ± 3.6 psi (0.25 bar).
- True pressure lies between 116.4 and 123.6 psi (8.03 to 8.52 bar), or ± 3.0% of the reading.
Pressure gauge accuracy calculator
Enter the range, the ASME grade, the reading and any temperature coefficient from the datasheet. It returns the zone, the pressure gauge accuracy band in psi and bar, and the band as a percent of reading.
Pressure gauge accuracy band calculator
What range should a pressure gauge be?
Pick a range so normal operating pressure sits near the middle of the dial, which usually means full scale at about twice normal pressure. ASME B40.100 guidance, as gauge makers summarize it, is to keep normal pressure near mid scale.
For the wider set of codes behind gauges, transmitters and valves, see this roundup of instrumentation and control standards.
Processing magazine describes the same habit: typical operating pressure at about half of full scale guards against over ranging during upsets. Running higher on the dial improves pressure gauge accuracy as a share of reading, but it raises the risk of overpressure.
Range rules for pressure gauge accuracy
- The WIKA 332.54 datasheet allows steady pressure up to full scale, fluctuating pressure up to 0.9 × full scale and short time pressure up to 1.3 × full scale.
- For Grades A and B, the tighter middle half band applies only from 25% to 75% of span: 50 to 150 psi (3.4 to 10.3 bar) on a 0 to 200 psi dial.
- If normal pressure is 120 psi and upsets reach 150 psi, 0 to 200 psi works; 0 to 160 psi puts upsets in the upper quarter.
Our post on the types of pressure gauges covers Bourdon, diaphragm and capsule designs.
Why does a pressure gauge lose accuracy?
Pressure gauge accuracy drifts from overpressure, pulsation, vibration, temperature swings, corrosion, clogging and rough handling. WIKA lists these among the most common reasons gauges fail in service.
Overpressure is the usual cause of a zero shift, because it strains the Bourdon tube past its elastic limit and leaves a permanent set. Ashcroft's selection guidance makes the same point: running near the top of the range can cause metal fatigue and drift in pressure gauge accuracy.
Temperature changes the spring rate of the element, which is why WIKA adds 0.4% of span for each 18 °F (10 K) away from 68 °F (20 °C). Our article on how ambient conditions affect instrument accuracy covers hot pump rooms and outdoor racks.
How do you troubleshoot poor pressure gauge accuracy?
Start at zero with the gauge vented, then compare against a better reference at several points going up and coming down. The pattern of the error points to the cause faster than any single reading.
| Symptom | Likely cause | Fix |
|---|---|---|
| Pointer off zero when vented | Overpressure, shock or a temperature shift | Rezero if the design allows; replace if the tube has taken a set |
| Pointer flutters | Pump pulsation or line vibration | Add a snubber from gauge accessories or use a liquid filled case |
| Up and down readings differ | Friction or a worn movement (hysteresis) | Recalibrate; replace if the gap is wider than the pressure gauge accuracy grade |
| Error grows with pressure | Span shift from fatigue or a bent link | Recalibrate span; move normal pressure toward mid scale |
| Reading sticks or lags | Clogged socket from solids or polymer | Fit a diaphragm seal and flush the connection |
| Wrong only on hot days | Temperature error on top of grade error | Add the coefficient to the pressure gauge accuracy budget, or move the gauge |
Test against a reference far better than the gauge, such as a Grade 4A test gauge or one of our pressure calibrators. Beamex also recommends exercising the gauge at maximum pressure for about a minute, then venting, two or three times before you record readings.

Field tips for keeping pressure gauge accuracy
- On liquid service, mount the gauge level with the tap or correct for head. Beamex shows that 1 m (3.3 ft) of water adds about 1.42 psi (98 mbar).
- Tag each gauge with its pressure gauge accuracy grade, range and last calibration date.
Pros and cons of higher pressure gauge accuracy grades
A tighter grade gives a smaller band, but it costs more and is often more delicate. Match pressure gauge accuracy to what the reading is used for, not to the best number in the catalog.
Pros
- Smaller psi band for setting relief valves and regulators
- Grades 4A to 1A hold one band over the whole dial
- Fewer disputes with the pressure transmitter on the same line
Cons
- Higher price and longer lead time
- Tight pressure gauge accuracy means little on the wrong range
- Test grade gauges suffer more from shock and pulsation
Myth vs fact: pressure gauge accuracy
Most mistakes come from reading the grade as a share of the reading or ignoring where the pointer sits. These four myths show up on almost every plant walk.
A 1% gauge reads within 1% of whatever it shows.
Pressure gauge accuracy is a percent of span. A 1% gauge on 0 to 100 psi can be off 1 psi at 10 psi, which is 10% of that reading.
A bigger range is always safer.
An oversized range widens the psi band and pushes the reading into the lower quarter, where Grades A and B are weakest.
The grade on the dial covers any temperature.
Pressure gauge accuracy grades apply at reference conditions. WIKA's 332.54 adds 0.4% of span per 18 °F (10 K) away from 68 °F (20 °C).
If the gauge reads zero, it is still accurate.
A gauge can hold zero and still have span or hysteresis error. Only a multi point test proves pressure gauge accuracy.
Industrial applications of pressure gauge accuracy grades
Plants choose the grade by what the reading drives: a local check, a setpoint or a test. In four common services, pressure gauge accuracy works out like this.
Refinery
Pump discharge gauges on hot units are often Grade 1A with a range near twice normal pressure. Technicians add the temperature error to the pressure gauge accuracy band when the gauge sits beside a hot line.
Water and wastewater plant
Raw water and sludge pumps pulse and carry solids, so gauges get snubbers or diaphragm seals to protect pressure gauge accuracy.
HVAC
Chilled and condenser water loops use gauges to read pump head and coil pressure drop. Differential pressure gauges avoid stacking two separate pressure gauge accuracy bands.
Chemical plant
Relief valve test benches use Grade 3A or 4A test gauges as references. Process gauges on corrosive service get seals to keep pressure gauge accuracy in grade.

How often should you check pressure gauge accuracy?
Most plants check process gauges once a year, and sooner on gauges that see pulsation, heat or overpressure. Fluke notes that most uncertainty analyses assume annual recalibration and points to ILAC G24 for setting intervals.
Shorten the interval when pressure gauge accuracy comes back out of grade, and stretch it only after several checks pass with margin. For process dial gauges built to ASME grades, see our pressure gauges page.
Frequently asked questions
What is the standard for pressure gauge accuracy in the US?
ASME B40.100 sets pressure gauge accuracy grades for US dial gauges. Grades 4A, 3A, 2A and 1A hold 0.1%, 0.25%, 0.5% and 1% of span across the dial. Grades A and B allow 1% and 2% in the middle half, with 2% and 3% in the outer quarters.
Is ASME Grade A or Grade 1A more accurate?
Grade 1A is the better of the two. It holds 1% of span over the whole dial, while Grade A holds 1% only in the middle half and 2% in the first and last quarter. If your reading may fall below 25% or above 75% of span, Grade 1A keeps pressure gauge accuracy tight.
How do I convert a pressure gauge accuracy grade into psi?
Multiply the grade percentage by the span in psi. A 0 to 200 psi gauge at 1% gives 2 psi (0.14 bar), and at 2% gives 4 psi (0.28 bar). Use the middle half value for Grades A and B only when the reading sits between 25% and 75% of span.
Why should normal pressure sit near the middle of the gauge scale?
Mid scale keeps the reading in the tightest zone for Grades A and B and leaves room for upsets. It also keeps the Bourdon tube well below its elastic limit, which limits fatigue and set. Processing magazine notes that operating at about half of full scale is common practice.
Does temperature change pressure gauge accuracy on a dial gauge?
Yes. Grades apply at reference conditions, and heat or cold changes the stiffness of the sensing element. WIKA's 332.54 datasheet adds 0.4% of span per 18 °F (10 K) away from 68 °F (20 °C), so a 0 to 200 psi gauge at 104 °F gains 1.6 psi of extra error.
Can a liquid filled gauge help readings on a pulsing pump?
A liquid fill damps pointer flutter, so the reading is easier to see and the movement wears less. WIKA notes that glycerin fill dampens vibration and pointer flutter in hydraulic service. It does not change the grade, so add a snubber where pulsation is severe and recheck after upsets.